Q:
How do I calibrate the meter of my Universal Audio LA-610?
A:
To calibrate your meter for the LA-610 follow the steps below.
1) Turn the LA-610 on and warm up for appx. 5 minutes. 2) Turn the GAIN knob to 0 3) Turn the Compressor section (right side) Peak Reduction control full off. 4) Adjust the GR Zero Set trim pot so the meter reads 0 dB. This is the small screw found directly underneath the UA logo, on the right side beneath the Meter window. Turn this small knob until the meter reads 0 db. These steps can also be found on page 6 of the LA-610 manual. |